Professional Testing Instruments Manufacturer
National consultation hotline: +86 021-64208466

Microflow Porometer

The PMI Microflow Porometer is a Capillary Flow Porometer with the ability to measure pore structure

Email: info@standard-groups.com

Content

The PMI Microflow Porometer is a Capillary Flow Porometer with the ability to measure pore structure and permeability in nearly impermeable materials. This instrument is user-friendly and fully automated. The Microflow Porometer is an asset in quality control and R&D environments.
 
Operation Principle
 In this instrument a pressure transducer is located on the downstream side of the sample and the time rate of pressure change in the known volume of the down stream side of the sample chamber is used to compute very small flow rates. The flow rate through dry samples is measured according to the differential pressure. Under the increasing pressure, the gas flows into the back of the completely wetted samples in the chamber. When the pressure increases enough to overcome the point where the capillary of the liquid works in the hole (maximum hole) , that point is the bubbling point. After confirming the bubbling point, increase the pressure and measure the flow until all the holes are empty and the sample is dry.
 
Application
 The microflow porometer is mainly used to determine the characteristics of low permeability membrane (gas separation membrane). The instrument is suitable for packaging, rubber abd special gas manufacturing companies.
Applied industries:
Automobile manufacturing      Battery separator  Ceramics  
Filtration   Geofabrics    
Nonwoven     Paper         
Pharmacy      Textiles
 
Features
· Samples permitting very low flow rate can be tested
· Very low permeability can be measured
· Fully Automated
· Windows based simple operation
· Provide other optional features
 
Technical Parameters

ItemsParameters
Pore Size Range0.013 - 500 microns
Permeability Range1x 10- 1x 10- (microflow in cc/sec/m/torr)
Sample size0.5” - 2.5” diameter
Pressure transducer range0~500psi
Resolution1 in 20,000
Accuracy0.15% of reading
Pressurizing gasClean, dry, and compressed air or nonflammable and non-corrosive gas
Mass flow transducer range10cm/minute~500,000cm/minute
Power requirements110/220VAC,50/60Hz (Others Available)
Dimension30” H x 19” W x 18.5” D
Weight100lbs

image.png

Disclaimer: The above content is for reference and communication only among industry insiders, and does not guarantee its accuracy or completeness. According to relevant laws and regulations and the regulations of this website, units or individuals who purchase related items should obtain valid qualifications and qualification conditions.