**Scope of Application**Determines the liquid loss performance of nonwoven fabrics. Place a standard
Email: info@standard-groups.com
**Scope of Application**
Determines the liquid loss performance of nonwoven fabrics. Place a standard absorbent medium beneath the test specimen, position the combined sample on an inclined platform, and measure the liquid penetration through the nonwoven fabric into the absorbent medium when a specific volume of artificial urine flows downward. Calculate the liquid loss performance by weighing the absorbent medium before and after testing.
**Compliance Standards**
Edana 152.0-99; ISO 9073-11
**Technical Specifications**
1. Test bench features two black reference lines:
- Low line: Located 3±0.2 mm above the bench end (absorbent medium position)
- High line: Approximately 25 mm below specimen top (aligns with drainage tube center)
2. Bench inclination: 25°
3. Clamp assembly:
- Adjustable device for fixing specimens symmetrically at (140±0.2) mm from reference line
4. Central positioning mechanism
5. Absorbent pad support frame at specimen lower edge
6. Glass tube specifications:
- Inner diameter: 5 mm
7. Ring-shaped base plate
8. Drip feeding system:
- Delivers (25±0.5) g test liquid continuously through glass tube in (4±0.1) seconds
Disclaimer: The above content is for reference and communication only among industry insiders, and does not guarantee its accuracy or completeness. According to relevant laws and regulations and the regulations of this website, units or individuals who purchase related items should obtain valid qualifications and qualification conditions.
Company Phone
+86-21-6420 0566
Working hours
Monday to Friday
Mobile phone:
13816217984
Email:
info@qinsun-lab.com